GDZD-601 Automatic Multi-Function Insulation Oil Shaker
1. 240*128 LCD display.
2. Small size, light weight and low noise.
3. Easy operation, control temperature is accurate.
1. Temperature measurement range: 0~120℃
2. Temperature control range: Room temperature to 110℃
3. Temperature control method: Digital PID auto control
4. Temperature control accuracy: Room temperature to 50℃, ±0.2℃; 50℃~100℃,±0.3℃; 100~110℃, ±0.5℃
5. Shaking frequency: 275±3 times/minute
6. Shaking amplitude: 35mm
7. Quantity of samples: Injector 100ml *8pcs; conical flask 250ml *4pcsA
8. Working modes:
Chromatography degassing: Constant temperature at 50℃, shaking for 20 mins, static for 10 mins.
Water-soluble acid: Constant temperature at 75℃, shaking for 5 mins, static for 0 minute.
User defined: Constant temperature 0 to 110℃ at any setting, shaking 0-99 mins at any setting, static 0~99 minutes at any setting. Start and stop by hand
Voltage: AC220V±20%
Frequency: 50Hz±5%
Max. power consumption: 800W
Environment temperature: 10℃-40℃
Environment humidity: ≤85%
Dimension: 500 X 350 X 400mm
Weight: 35 kg