Open Cup Flash Point Tester Overview and Technical Specifications

Open Cup Flash Point Tester Overview and Technical Specifications

The flash point, also known as the ignition point, is the lowest temperature at which a mixture of the material or product with the surrounding air comes into contact with a flame and causes a flash and immediate combustion. It is an important indicator for safety protection during the storage, transportation, and use of materials or products. In many cases, the open flash point tester plays a significant role. Now, let’s learn about the overview and technical indicators of the open flash point tester together with Guodian Xigao.
I. Overview
The  HZKK-208 Open Flash Point Tester is a flash point testing instrument that complies with standards GB267-2008 and GB3536-2008 as well as ISO2592, and features a unique self-defined standard for conducting tests. It employs microprocessor technology, has a large-screen LCD display in Chinese, and automatically prints the test results. It has a high degree of automation and strong human-computer interaction capabilities. This instrument features automatic height adjustment, automatic temperature rise, switching, automatic fire sweeping and ignition, two options of electric and gas ignition, automatic end detection and result printing, and automatic air cooling. The entire testing process is fully automated. It can store and delete test results at will.

II. Basic Technical Specifications
Display mode: Large screen LCD Chinese characters display
Measurement range: Room temperature to 400℃
Measurement accuracy: 0.1℃
Measurement repeatability: Error ≤ ±2℃ at temperatures below 150℃
At 150℃, the error is ±4℃.
Cooling method: Forced air cooling
Operating temperature: 10 – 55℃
Use humidity: 30 – 80%
Power supply: AC 220V ± 5%, 50Hz
Power consumption: 700VA
Dimensions: 455 × 365 × 300 (mm)


Post time: Jun-02-2026