The usage method of the battery charge and discharge tester:
Preparation before testing: Connect the battery to the battery charge and discharge tester correctly and set the test parameters.
Charging test: Charge the battery through the charging function, and test parameters such as charging time and charging current. Record the data after the test is completed.
Discharge test: Discharge the battery through the discharge function to test parameters such as discharge time and discharge current. Record the data after the test is completed.
Sort out the data: Based on the test results, sort out the data and analyze the parameters of the battery such as electrical energy and capacity.
Diagnose the state of the battery: Based on the test results and data analysis, determine the state of the battery.
Select an appropriate testing environment: During the testing process, it is necessary to choose an appropriate environment to avoid factors such as electromagnetic interference and strong light that may affect the test results.
Precautions for Battery Charge and Discharge Tester:
Please read the instruction manual carefully before using the battery charge and discharge tester, operate it in the correct way and keep the manual properly for future reference.
The paragraphs marked with “Attention” in the instruction manual are general warnings, and those marked with “Danger” are important warnings. Please read them carefully and pay close attention.
Ensure that the storage and usage environment of the battery charge and discharge tester is clean and safe. Before use, carefully check if the machine is in good condition.
For safety reasons, the grounding terminal of this machine must be reliably grounded before use.
It is strictly forbidden to operate with power on when connecting or disconnecting wires.
When the battery charge and discharge tester is connected to the DC power supply and the battery pack, be sure not to reverse the positive and negative poles. The connection must be firm and free of looseness before the machine can be turned on.
The battery charge and discharge tester should operate in an environment free of corrosive, acidic, alkaline, flammable, explosive and other dangerous substances.
The battery charge and discharge tester should be operated in a clean, tidy and well-ventilated indoor open space. There should be no obstacles within 4 to 5 meters in front of the air inlet and outlet of the equipment.
If working outdoors is necessary, it is essential to avoid doing so under adverse conditions such as direct sunlight, rain splashing, and against the wind to prevent damage to this machine.
When the battery charge and discharge tester is in operation, do not touch the internal components of the machine to avoid electric shock.
After the work is completed, please let the fan continue to run for about 5 minutes to dissipate the residual heat inside the case. Then, turn off the power of this machine. Finally, remove all the connections and restore the original state before the work.