China High Quality Meter test kit Supplier – GDYB-S3 Three Phase Energy Meter Test System (3 positions) – HV Hipot
China High Quality Meter test kit Supplier – GDYB-S3 Three Phase Energy Meter Test System (3 positions) – HV Hipot Detail:
Technical Specification
System accuracy: 0.1%;
Three phase source parameter:
a. Output voltage:
3*57.7VY 3*100V ∆ ;
3*220V Y 3*380V ∆;
Adjustable range 0~120% continuously adjustable
Adjustment fineness: better than 0.01%
b. Output current: 25mA, 0.1A, 1A, 5A, 25A,100A;
Adjustable range 0~120% continuously adjustable
Adjustment fineness: better than 0.01%
Mininum output current: 1mA (accuracy: <5%)
c. Output power: (voltage circuit) 3 * 50VA
(current circuit) 3 * 150VA
d. Output voltage, current, power stability: ≤0.02%/120S
e. Distortion degree of output voltage and current waveform: ≤0.5%
f. Symmetry of three phase angles: between 120°±0.3°
g. Frequency range: 45Hz~ 65Hz, step: 0.01Hz
h. Phase angle range: -180°~ 180°, step: 0.01°
i. Indicating metter accuracy: Voltmeter, ammeter: ±0.01%
Phase meter: ±0.01°
Frequency meter: ±0.01Hz
Minimum Output current: 1mA; accuracy: ≤5%;
Output load capacity (20 meters)
Voltage output: 800VA;
Current output: 2500VA;
Voltage & current adjustment
Adjustable range: (0%~120%) x measuring range;
Adjustable fineness: 10%, 1%, 0.1%, 0.01%;
Phase adjustment:
Adjustable range: 0°~360°;
Adjustable fineness: 10°, 1°, 0.1°, 0.01°;
Frequency adjustment:
Adjustable range: 45Hz~65Hz;
Adjustable fineness: 1Hz, 0.1Hz, 0.01Hz;
Output power stability: ≤0.05%/120S;
Output voltage & current distortion:
Voltage: ≤0.5%;
Current: ≤0.5%;
Harmonic settings:
Harmonic number: 2~21 times;
Harmonic content: current 2~21 times ≤40% (relative to fundamental wave);
voltage 2~21 times ≤40% (relative to fundamental wave);
Harmonic phase: 0°~359.99°;
Monitoring instrument:
Standard meter display accuracy: Voltage: 0.01%;
Current: 0.01%;
Power: 0.01%;
Phase: 0.01;
Frequency: 0.01;
Insulated 1:1 current transformer parameter:
Accuracy: 0.01%;
Output power: 60VA;
Current measuring range: 1mA~120A (1:1);
Main performance index
a) Accuracy: 0.01% (COS∮=0.5C…1…0.5L);
b) Rated frequency: 50Hz/60Hz±5%;
c) Primary current: 3×0.001-120A;
d) Secondary current: 3×0.001-120A;
e) Output load (each phase):
0.5W (output range: 0.001-1A);
0.7V divided by current value (output range: 1-120A);
f) Output load voltage: Max. 0.5V (output current Max. 120A);
Current transformer dimension: 265 x 200 x 240 mm (L x W x H);
Weight: 12kg;
Dimension:
Control source: 610 x 800 x 1920mm (L x W x H);
Meter rack: 2350 x 750 x 1850mm (L x W x H);
Product detail pictures:
Related Product Guide:
We've been proud with the significant shopper fulfillment and wide acceptance due to our persistent pursuit of top of the range both of those on solution and repair for China High Quality Meter test kit Supplier – GDYB-S3 Three Phase Energy Meter Test System (3 positions) – HV Hipot , The product will supply to all over the world, such as: Houston, Ecuador, Detroit, With more and more Chinese products and solutions around the world, our international business is developing rapidly and economic indicators big increase year by year. We have enough confidence to supply you both better solutions and service, because we've been more and more powerful, specialist and experience in domestic and international.