Technical Specification
System accuracy: 0.1%;
Three phase source parameter:
a. Output voltage:
3*57.7VY 3*100V ∆ ;
3*220V Y 3*380V ∆;
Adjustable range 0~120% continuously adjustable
Adjustment fineness: better than 0.01%
b. Output current: 25mA, 0.1A, 1A, 5A, 25A,100A;
Adjustable range 0~120% continuously adjustable
Adjustment fineness: better than 0.01%
Mininum output current: 1mA (accuracy: <5%)
c. Output power: (voltage circuit) 3 * 50VA
(current circuit) 3 * 150VA
d. Output voltage, current, power stability: ≤0.02%/120S
e. Distortion degree of output voltage and current waveform: ≤0.5%
f. Symmetry of three phase angles: between 120°±0.3°
g. Frequency range: 45Hz~ 65Hz, step: 0.01Hz
h. Phase angle range: -180°~ 180°, step: 0.01°
i. Indicating metter accuracy: Voltmeter, ammeter: ±0.01%
Phase meter: ±0.01°
Frequency meter: ±0.01Hz
Minimum Output current: 1mA; accuracy: ≤5%;
Output load capacity (20 meters)
Voltage output: 800VA;
Current output: 2500VA;
Voltage & current adjustment
Adjustable range: (0%~120%) x measuring range;
Adjustable fineness: 10%, 1%, 0.1%, 0.01%;
Phase adjustment:
Adjustable range: 0°~360°;
Adjustable fineness: 10°, 1°, 0.1°, 0.01°;
Frequency adjustment:
Adjustable range: 45Hz~65Hz;
Adjustable fineness: 1Hz, 0.1Hz, 0.01Hz;
Output power stability: ≤0.05%/120S;
Output voltage & current distortion:
Voltage: ≤0.5%;
Current: ≤0.5%;
Harmonic settings:
Harmonic number: 2~21 times;
Harmonic content: current 2~21 times ≤40% (relative to fundamental wave);
voltage 2~21 times ≤40% (relative to fundamental wave);
Harmonic phase: 0°~359.99°;
Monitoring instrument:
Standard meter display accuracy: Voltage: 0.01%;
Current: 0.01%;
Power: 0.01%;
Phase: 0.01;
Frequency: 0.01;
Insulated 1:1 current transformer parameter:
Accuracy: 0.01%;
Output power: 60VA;
Current measuring range: 1mA~120A (1:1);
Main performance index
a) Accuracy: 0.01% (COS∮=0.5C…1…0.5L);
b) Rated frequency: 50Hz/60Hz±5%;
c) Primary current: 3×0.001-120A;
d) Secondary current: 3×0.001-120A;
e) Output load (each phase):
0.5W (output range: 0.001-1A);
0.7V divided by current value (output range: 1-120A);
f) Output load voltage: Max. 0.5V (output current Max. 120A);
Current transformer dimension: 265 x 200 x 240 mm (L x W x H);
Weight: 12kg;
Dimension:
Control source: 610 x 800 x 1920mm (L x W x H);
Meter rack: 2350 x 750 x 1850mm (L x W x H);
Product detail pictures:
Related Product Guide:
To be a result of ours specialty and repair consciousness, our corporation has won an excellent reputation amongst customers all around the entire world for OEM Best energy meter calibration equipment Factories – GDYB-S3 Three Phase Energy Meter Test System (3 positions) – HV Hipot , The product will supply to all over the world, such as: Belgium, USA, luzern, Our company has built stable business relationships with many well-known domestic companies as well as oversea customers. With the goal of providing high quality products to customers at low cots, we've been committed to improving its capacities in research, development, manufacturing and management. We have honored to receive recognition from our customers. Till now we have now passed ISO9001 in 2005 and ISO/TS16949 in 2008. Enterprises of "quality of survival, the credibility of development" for the purpose, sincerely welcome domestic and foreign businessmen to visit to discuss cooperation.